USPSIMULTANEOUS DECONTAMINATION AND STERILIZATION

Nov. 19, 2025

Surfaces to be sterilized should use a higher challenge population (106 spores per unit) of a resistant BI with the expectation of no recoverable microorganisms post-process. This provides an estimated 8–9 log reduction in proximity to the BI.

 

The combination processes would, simliar to other decontamination and sterilization processes, be performed on previously cleaned surfaces to ensure full exposure to the sporicidal agent. Consistent with these processes, the pretreatment bioburden is evaluated to confirm that it is less resistant to the chosen agent and less populous than the BI microorganism chosen, affording greater confidence in the outcome. Under these circumstances, the surfaces confirmed with the higher challenge population are rendered “sterile”, whereas the remaining surfaces would be considered “decontaminated”.

 

The target surfaces and items in systems decontaminated in this manner must be fully exposed to the decontaminating agent. Effectiveness of the disinfection process requires sufficient concentration of the sporicidal agent, as well as the proper range of temperature and relative humidity. Systems treated in this manner must maintain a positive pressure differential to the external environment, and must be supplied with high-efficiency particulate air (HEPA)-filtered air, providing International Organization for Standardization (ISO) 5 conditions internally. This will prevent the ingress of microbial contamination after the disinfection process has been completed.

 

VALIDATION

The cycle development and validation approaches for these disinfection processes, described in 〈1229.7〉 and 〈1229.11〉, apply to their use in combined decontamination and sterilization with only minor changes. The parameters to be established include the following: the quantity of sporicide used (derived from concentration measurement or amount delivered), relative humidity, and temperature. The BIs used for confirmation of process effectiveness are the same as those indicated for sterilization using the same agent. The D-value of the challenge microorganism can be used in conjunction with gas sterilization processes (see 〈1229.7〉), whereas vapor processes are subject to the same limitations described for vapor sterilization, and require empirical assessment (see 〈1229.11〉).

The following activities are defined for a combined decontamination and sterilization process.

 

Equipment Qualification

The equipment qualification for combination disinfection systems generally follows that of sterilization processes to confirm that the equipment has been properly installed and operates as intended.

 

Empty Chamber Parameter Distribution

The multipoint concentration measurement can be utilized for gas processes, but is not useful with vapor phase delivery systems. Humidity and temperature measurements, along with chemical indicators, can provide a limited indication of agent distribution. BIs are not required in the evaluation of the empty chamber.

 

Load Item Mapping

For surface treatment, internal mapping of load items is not required. Where items subject to interaction with the agent are present in the system, the lack of adverse impact on them post-decontamination should be evaluated.

Load Mapping

The effects of internal load size and patterns on the process should be assessed.

 

Biological Indicators

The use of multiple BIs at each test location is recommended to better support process lethality. The preference is to use BIs from the same lot with different populations, depending upon the target item on which they are positioned. The details of BI preparation and use are described in Biological Indicators for Sterilization 〈1229.5〉.

 

Process Confirmation and Microbiological Challenge

The goal of the validation activity is the confirmation of acceptable process parameters and inactivation of the microbial challenge. As noted previously, different BI populations are used to demonstrate decontamination and sterilization simultaneously. The proof of cycle efficacy is provided in replicate studies in which the BIs are killed and chemical and/or physical measurements are utilized.

 

ROUTINE PROCESS CONTROL

The combined decontamination and sterilization process is subject to overall controls that maintain it in a validated state. The practices outlined in Sterilization of Compendial Articles 〈1229〉 describe the general requirements appropriate for all sterilization systems. The essential practices required to maintain validated status include calibration, physical measurements, use of BIs, use of physical or chemical integrators and indicators, ongoing process controls, change controls, preventive maintenance, periodic reassessment, and training.

 

REFERENCES

1.Parenteral Drug Association. PDA Technical Report No. 28 Revised. Process simulation testing for sterile bulk pharmaceutical chemicals. PDA J. Pharm. Sci. Technol. 2006;Supplement 60:S-2.

2.US Food and Drug Administration. Guidance for industry. Sterile drug products produced by aseptic processing–current good manufacturing practice. Rockville, MD: US Food and Drug Administration; October 2004. https://www.fda.gov/regulatory-information/search-fda-guidance-documents/sterile-drug-products-produced-aseptic-processing-current-good-manufacturing-practice.

3.Pharmaceutical Inspection Convention. Pharmaceutical Inspection Co-Operation Scheme. Recommendation. Isolators used for aseptic processing and sterility testing, PI-014-2, July 2004.

4.Parenteral Drug Association. PDA Technical Report No. 34. Design and validation of isolator systems for the manufacturing and testing of health care products. PDA J. Pharm. Sci. Technol. 2001;55(5):1–23.▲ (USP 1-May-2023)

 

NOTE:Source from USP,The above content is for reference only. For more information, please click the original text:https://online.uspnf.com/uspnf/document/2_GUID-0D3B6D1D-D28C-4AF2-AC72-4B19C4599275_10101_en-US?source=TOC

 