INTRODUCTION
Steam sterilization of aqueous liquids (including both suspensions and emulsions with mixing), also known as sterilization of nonporous loads, is the method of choice for aqueous parenteral products, in-process aqueous liquids, laboratory media, and biological waste materials. This type of sterilization is accomplished primarily in closed containers. During steam sterilization by direct contact (also called steam sterilization of parts, hard goods, or porous items) the steam in the chamber directly contacts the surface of load items to effect sterilization (see Steam Sterilization by Direct Contact 〈1229.1〉). In contrast, sterilization of liquids in containers is accomplished by application of heat to the container, heating of the container wall, and ultimately heating of the internal liquid volume. This can be accomplished using steam, superheated water, and air in various combinations. Some aqueous liquids are susceptible to over-processing that could render them unfit for their intended use. Manufacturers should consider the influence of these differences when they establish a suitable process.
During the sterilization of liquid-filled containers, differential pressures between the interior of the containers and the sterilization chamber may potentially impact container integrity. Air over-pressure is used to minimize the pressure differential between the container and the sterilizer to protect the integrity of the container, especially prefilled syringes and plastic containers. Before sterilization of product containers, manufacturers should consider the potential adverse consequences of excess heat on the materials. In order to ensure sterility as well as functionality, the process definition and validation method used must incorporate both lower and upper temperature and time limits on the process conditions.
When the overkill method can be used for sterilization of sealed liquid containers, it is the preferred method and is described in 〈1229.1〉. When product quality attributes can be impaired by excessive heat, the sterilization process should use less time or a lower temperature to minimize the adverse effect on the materials. Sterilization time–temperature or F0 conditions (F0 is defined as the equivalent sterilization time relative to a base temperature of 121°) include both lower (sterility-related) and upper (stability-related) limits.1 Manufacturers commonly employ the bioburden/biological indicator (BB/BI) or bioburden methods when constraints on the material's ability to withstand the process require the use of less aggressive conditions. This approach requires appropriate controls on presterilization bioburden and/or product-related D-values in conjunction with bioindicators of lower spore count or resistance to ensure sterilization.
Terminal Sterilization of Products
The maintenance of product attributes may require the use of sterilizing conditions that are less aggressive and sterilization equipment, cycles, and validation methods adapted to these more restricted circumstances. The substantial variations in equipment designs and methods for terminal sterilization preclude a singular description of a typical cycle. All terminal sterilizers heat the load, but they accomplish this in varying ways: saturated steam; steam–air mixtures, steam–air–water mixtures, and superheated water. Air over-pressure for maintenance of container integrity and cooling containers and water for heating/cooling of the load may be present depending upon the autoclave size, throughput expectations, and container.
In-Process Fluids
In-process fluids are used for pH adjustment, dilution to a specified volume, lubrication, and other purposes. In many instances these liquids are sterilized in conjunction with items that must be sterilized by direct steam contact, and the sterilization process must ensure that all items are adequately sterilized.
Laboratory Media
Laboratory media often are sterilized in standard steam autoclaves with minor adaptations. Provision for slow exhaust (to reduce stress on container integrity and minimize boil over) and jacket cooling can help improve the basic steam sterilizer design and operation to better accommodate the materials. The sterilization process may be specific for media containers or a combination of both liquid-filled containers and hard goods. This process may resemble the methods used for terminally sterilized products (see above). The sterilization of laboratory media may entail the processing of a number of different containers that contain different materials. Manufacturers should be aware of the potential for under- and over-processing across the load and must consider container size, container contents, and position. When liquid-filled containers are combined in the same load with hard goods, manufacturers must consider the unique concerns of each to ensure all items are properly sterilized. Because laboratory media are considered self-indicating with respect to sterility, the use of internal biological indicators during validation is not required.
Biowaste Sterilization
The sterilization of biowaste in sealed containers from laboratory or production use is similar to parts sterilization. The process is defined to ensure a minimum time–temperature exposure or attainment of a specified F0 value throughout all items of the load. Depending on the potential contaminants present, the autoclave design may incorporate condensate collection/sterilization or sterilizable exhaust filters to ensure that pathogens are adequately contained. Because the objective of biowaste sterilization is to render the materials safe for contact and disposal, the overkill method described in 〈1229.1〉 is employed.